Manufacturer: BÄHR-Thermoanalyse GmbH (Germany)
Equipment model: DIL805A
Location: Room 102 of Science and Technology Building
Person-in-charge: Guo Jin
Tel.: 62332598-6103
Main functions:
Used for phase transition test and study of metal material. The temperature range of the corresponding structure can be obtained by measuring the change in length of the specimen during heating and cooling.
1) Determination of phase transition point during heating, such as determination of AC1 and AC3 of steel.
2) Determination of phase transition point during continuous cooling, to obtain a CCT curve.
3) Determination of phase transition point during the constant temperature process after rapid cooling, to obtain a TTT curve.
4) Determination of coefficient of linear expansion during heating and cooling.
5) Simulation of heat treatment process and welding process.
Main technical parameters
Specimen range: conductive solid specimen
Specimen size: φ4 × 10 mm
Temperature range: room temperature ~ 1,250°C
Heating method: induction heating
Ambient atmosphere: vacuum or argon atmosphere, vacuum degree: 5 × 10-4 mbar
Cooling medium: nitrogen
Cooling rate: up to 200°C/s
Accuracy: 0.05 μm/0.05°C
Laboratory appointment:
http://202.204.49.249/Equipment/Show?Id=""9f06cd9d-ba92-4748-a96a-22151fe0b28e